?All wetted parts made of Series 300 stainless steel ?Provided in a rugged NEMA 4 enclosure ?Perfect for harsh applications such as chillers, boilers and refrigerant systems 美國controller差壓變送器560系列
詳細介紹
美國controller差壓變送器560系列
Series 560 Wet-to-Wet Pressure Transducer
Features and Applications
All wetted parts made of Series 300 stainless steel
Provided in a rugged NEMA 4 enclosure
Perfect for harsh applications such as chillers, boilers and refrigerant systems
Available in both gauge and differential versions
Customizable for any pressure range from 0/20PSI to 0/500PSI
One of the most cost-effective wet-to-wet transducers on the market
Specifications
Accuracy
+/-0.5% FSO
Stability:
+/-0.5% FSO/yr
Thermal Effects:
+/-0.02% FSO/°c
Overpressure:
2 times FSP or 700PSI whichever is less
FSO Pressure Range:
0/20psi up to 0/500PSI
Compensated Range:
10° to 50°C
Media:
Any media, wet or dry, compatible with series 300 stainless steel
Operating Humidity:
NEMA 4 levels with proper conduit connections
Operating Temp:
-25° to 70°C
Input Supply**:
24VDC/24VAC
Supply Current:
<35ma<>
Load Resistance:
2KΩ minimum on voltage out, 700Ω, maximum loop resistance on 4-20mA output
Output Signal:
0-5VDC, 0-10VDC or 4-20mA
Electrical Connections:
Internal screw terminals
Pressure Connections:
1/8″ NPT Female
Enclosure:
NEMA 4 enclosure
Dimensions:
Approx. 5.12″ x 3.70″ x 2.25″ (13.00cm x 9.40cm x 5.72cm)
* Includes non-linearity hysteresis and non-repeatability at a fixed temperature
Package Drawing
How to Order
A typical order number consists of the model number, pressure range, pressure unit, output and mounting connection.
Model
Type
Output Signal
FSO Pressure Range
Calibration Pressure
560
D = Differential or Gage
0-5VDC
4…..20PSI to 50PSI
Customer Designated exact FSO
0-10VDC
6…..50PSI to 100PSI
4-20mA
7…..100PSI to 500PSI
Example: 560D 4-20mA Cal Pressure = 0-30 PSI (Model number 560, differential type unit, output signal of 4-20mA, pressure range up to 500PSI, calibrated to 0-30PSI FSO).